The pump should be cleaned regularly every 3 months. Please do not power on when cleaning. Use 95% alcohol solution and wet paper towel to clean and disinfect when cleaning. Do not use corrosive chemical solution cleaners. After cleaning, wait for the machine to be completely dry before power on. It is forbidden to use high-pressure instruments for sterilization.
1. Please check whether the infusion tube is damaged or leaking, and then check whether the infusion tube is installed correctly and whether the bubble detector is dirty,
2. Turn on the machine to check whether the sensitivity parameter setting of the bubble detector is too high. If the parameter setting is too high, please adjust and lower the setting parameter. The specific parameter settings are adjusted as follows: after starting up, press the start key for a long time to enter the selection interface, press the password interface for 3 consecutive times, press the page turning key to enter the bubble sensitivity level interface after entering the password 018.0, press the down key to 120, and then press the page turning key to confirm OK. If the "bubble alarm" problem cannot be solved after the above operations are completed, please send the machine back to our company's after-sales service department, and we will help you solve this problem.
1. Check whether the preset quantity is correct;
2. Whether an uncalibrated specification or brand of infusion line is used, it can only be used after recalibration.
1. Syringe pump: correctly install the syringe on the machine, press and hold the start key while turning on the power switch, enter the password interface, enter the password 012.0, press the shuttle to enter, select the syringe brand and press the shuttle to confirm, then rotate to "OK" and press the shuttle to calibrate.
2. Infusion pump: long press the start key to enter the interface after starting up, then press the start key twice in succession to enter the password interface, press the start key to enter the calibration interface after entering the password 012.0, a. press the up and down key and select key to select the brand of infusion set, then press the select key to enter the next setting item, B. set the amount of liquid medicine to be calibrated (10ml),c. Set the number of drops of 1ml (20 drops) d. Calibrate the start of infusion. Drop the output liquid medicine into the measuring cup, check the difference between the actual liquid medicine volume in the measuring cup and the set value (10ml), input the difference value into the calibration result as the compensation value, then carry out the second calibration, debug 6 times in turn, and the machine shows that the calibration is completed.
1. Maintenance of the main engine: Use 95% alcohol solvent or detergent and wet paper towel to wipe and test the shell of the main engine and dry it. Do not let the water in the water chamber flow backward into the machine due to immersion or inclination of the water chamber;
2. Pipes and masks: Soak in warm boiled water with neutral detergent for about 30 minutes, then take it out and rinse it with water, dry it before use, and clean it at least once a week;
3. The filter sponge is cleaned once a week and replaced once a month.
1. First check whether the "automatic on/off" item in the parameter setting of the ventilator is set to "on". If the status is "on", it needs to be debugged to "off;
2. The mask may not have a good air leakage, the machine will start the self-protection program to automatically shut down;
If it is found that the motor is running abnormally, such as suddenly stopping, running too fast, loose coordination of the mobile phone, vibration, sudden noise, excessive temperature rise, etc., please do not continue to use it.
Before replacing the root canal file, it must be turned off. Failure to do so may result in accidental rotation of the file needle and personal injury.
When removing the root canal file, first press the gland of the mobile phone chuck, and then unplug the root canal file. Do not forcibly remove the root canal file, otherwise the chuck of the mobile phone chuck is easily damaged. For other details, please refer to the instruction manual.
It will not affect the normal use of the root canal preparation machine, and its speed and torque will not change.
Note: After each use is completed, no matter how much the battery power, can be placed on the charging base to charge, so as to use with the charge, will not affect the battery life. Please refer to the manual for other charging precautions.
undefined
1. Disconnect the corridor display bus;
2. The corridor display screen is broken;
3. The host power board is damaged;
4, the host motherboard corridor display output chip burned (317 or 324 chip burned)
1. Disconnect the corridor display bus;
2. The corridor display screen is broken;
3. The host power board is damaged;
4, the host motherboard corridor display output chip burned (317 or 324 chip burned)
1, may be stuck;
2. The cancellation of the bathroom extension needs to be turned off at the host end, and other positions cannot be turned off;